德国Elementar CHONS元素vario PYRO cube

发布:2016-10-04 13:59 | 来源:未知 | 浏览:

vario PYRO cube

  • 样品适应性广-一台仪器可满足高温燃烧的CNS模式和高温裂解OH模式测定
  • 市面唯一的零空白的O-IRMS分析元素分析仪
  • 专利的反吹技术,确保基质独立的可靠结果
  • 两仓的热裂解管,确保高精度和长寿命
  • 可选配高通量的液体自动进样器-同时满足CNS和OH模式
  • 简单易用,智能的软件自诊断功能
  • 专利的球阀进样,零背景空白

A reliable allrounder

Stable isotope laboratories have ever increasing demands for functionality and versatility. H and O isotopic determination in solids and liquids often requires pyrolysis at temperature of up to 1,500 °C. At the same time, laboratories need routine for carbon, nitrogen, and sulfur isotope ratio determination as well as elemental concentration analysis. In the past, laboratories have had to invest in two separate analyzers to cover this range of analysis; one for combustion and the other for pyrolysis.

The vario PYRO cube breaks new ground by combining an up to 1,500 °C pyrolysis analyzer with a fully functional elemental analyzer for carbon, nitrogen, and/or sulfur determination with separate combustion and reduction tubes.

Backflush technology for unbiased results

Oxygen isotope ratio determination of high nitrogen containing samples poses analytical challenges. The use of Elementar purge and trap technology results in complete baseline separation of the N2 from the CO peak and superior CO peak focussing. The system uses a special backflush system (patent applied for) to remove background bleed of the pyrolysis reactor from the CO peaks.

The vario PYRO cube is a member of Elementar's well renowned "cube family". The very small footprint is the result of the most advanced functional design in the market. Even the power supply for the 1,500 °C furnace is incorporated in the compact design of the vario PYRO cube with a footprint of only 42 x 55 cm.
 

Specifications

High temperature pyrolysis
quantitative sample pyrolysis at up to 1,500 °C . This is the prerequisite for 100% recovery also for samples difficult to pyrolyze. Equipped with one silicon carbide heater element and a standard reduction oven, it allows full flexibility without any compromise.

Low instrument background
Due to the novel reactor design with fully separated gas flow a very low background signal has been achieved.

Largest dynamic range of element concentrations and element ratio
Sample weight between 0.02 mg and 1 g with up to 6 mg O abs., from ppm to 100%

Purge&trap Chromatography
Separation of gaseous components on up to three specific columns for perfect separation without overlap and automatic optimization of the analysis time. A peak separation to the baseline, e.g. for CO2 and N2, can be reached even with a concentration ratio of 5000:1.

Backflush technology
Specifically equipped with novel backflush technology to guarantee full nitrogen-carbon monoxide separation for highly accurate analysis of oxygen.

Instrument control
with a standard PC for perfect results also for unknown or heavily variable sample matrices with integrated sample feeder for 120 samples. Possibility for remote service due to full digitalization of the instrument electronics

One basic instrument for all applications
No expensive special catalysts, easy ash removal from the pyrolysis tube and long lifetime of all components are the basis for a low cost of ownership

Clean, safe, fast
The analysis only takes a few minutes, no aggressive or toxic chemicals are used, Helium carrier gas

Few installation requirements
48 x 55 cm benchtop space, 1 power supply (230, 110, 100 V), and two gases are sufficient for maintenance free long time operation

Isotope analysis